ASTM G138 Standard Test Method for Calibration of a Spectroradiometer Using a Standard Source of Irradiance
ASTM International
Standard Test Method for Calibration of a Spectroradiometer Using a Standard Source of Irradiance
N G138
Annotation
This test method covers the calibration of spectroradiometers for the measurement of spectral irradiance using a standard of spectral irradiance that is traceable to a national metrological laboratory that has participated in intercomparisons of standards of spectral irradiance.
This method is not limited by the input optics of the spectroradiometric system. However, choice of input optics affects the overall uncertainty of the calibration.
This method is not limited by the type of monochromator or optical detector used in the spectroradiometer system. Parts of the method may not apply to determine which parts apply to the specific spectroradiometer being used. It is important that the choice of monochromator and detector be appropriate for the wavelength range of interest for the calibration. Though the method generally applies to photodiode array detector based systems, the user should note that these types of spectroradiometers often suffer from stray light problems and have limited dynamic range. Diode array spectroradiometers are not recommended for use in the ultraviolet range unless these specific problems are addressed.
The calibration described in this method employs the use of a standard of spectral irradiance. The standard of spectral irradiance must have known spectral irradiance values at given wavelengths for a specific input current and clearly defined measurement geometry. Uncertainties must also be known for the spectral irradiance values. The values assigned to this standard must be traceable to a national metrological laboratory that has participated in intercomparisons of standards of spectral irradiance. These standards may be obtained from a number of national standards laboratories and commercial laboratories. The spectral irradiance standards consist mainly of tungsten halogen lamps with coiled filaments enclosed in a quartz envelope, though other types of lamps are used. Standards can be obtained with calibration values covering all or part of the wavelength range from 200 to 4500 nm.
Скачать документ нельзя
Вы можете заказать документ
Международные и зарубежные стандарты (ASTM, ISO, ASME, API, DIN, BS и др.) не предоставляются в рамках данной услуги. Каждый стандарт приобретается платно с учетом лицензионной политики Разработчика.
Любые авторские документы, размещенные на сайте, представлены в соответствии с признанным в международной практике принципом «как есть». ООО «Информпроект Групп» не несет ответственности за правильность информации, изложенной в авторских документах.
или посмотрите возможности крупнейшей электронной библиотеки «Техэксперт» — более 25 000 000 документов!